FDA Brief

Recalls

When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.

5,412 recalls in the last 12 months. Newest first.

ReportedClassFirmProduct and reason
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0685-2026
Dec 3, 2025 Class II BioFire Diagnostics, LLC
Device
BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
FDA notice Z-0602-2026
Dec 3, 2025 Class II Medtronic Sofamor Danek USA Inc
Device
Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT PL SHORT 7MM. Model Number: 6068073. 2. SPACER 6068076 CATALYFT PL LONG 7MM, Model Numb…
Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be updated to mitigate risk of loss of lordosis to occur.
FDA notice Z-0580-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding ca…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0625-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0643-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0706-2026
Dec 3, 2025 Class II INTELERAD MEDICAL SYSTEMS INCORPORATED
Device
IntelePACS - InteleConnect / TechPortal
Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.
FDA notice Z-0723-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0690-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresp…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0612-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with correspondi…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0634-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0657-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0667-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresp…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0608-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with correspondi…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0631-2026
Dec 3, 2025 Class II Magellan Diagnostics, Inc.
Device
The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare¿ II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 …
Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
FDA notice Z-0584-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0650-2026
Dec 3, 2025 Class II Inpeco S.A.
Device
FlexLab (FLX); Version: FLX-217-40;
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
FDA notice Z-0596-2026
Dec 3, 2025 Class I Glycar SA Pty., Ltd.
Device
SJM Pericardial Patch, with EnCap Technology 9x14cm, Model Number REF C0914; PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE
The impacted lot may not meet the required tensile strength specification.
FDA notice Z-0531-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0674-2026
Dec 3, 2025 Class II Olympus Corporation of the Americas
Device
Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light sourc…
Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.
FDA notice Z-0579-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0641-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with correspondi…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0635-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0707-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresp…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0613-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
RATE FLOW set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with correspondi…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0710-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0663-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0665-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0652-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0704-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0666-2026
Dec 3, 2025 Class II DJO, LLC
Device
Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02
The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.
FDA notice Z-0722-2026
Dec 3, 2025 Class II Pivotal Health Solutions, Inc.
Device
Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
FDA notice Z-0587-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0694-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0658-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0681-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0699-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0668-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0672-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding cat…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0604-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0692-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0682-2026
Dec 3, 2025 Class II Foundation Medicine, Inc.
Device
FoundationOne Liquid CDx (F1LCDx), TECH-0009 version 6.0
Reports were sent to customers without the companion diagnostic (CDx) Claims Page. Claims pages were distributed via amended reports within 7 days of the issue occurrence.
FDA notice Z-0578-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0664-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresp…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0614-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0698-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with correspondi…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0638-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0677-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0654-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0661-2026
Dec 3, 2025 Class II B Braun Medical Inc
Device
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with cor…
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
FDA notice Z-0687-2026