Recalls
When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.
5,412 recalls in the last 12 months. Newest first.
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Aug 12 | Class III | Central Admixture Pharmacy Services Inc Drug |
Total Parental Nutrition - Pediatric PN Patient-Specific TPN Bag, (patient specific), Compound Volume 416.8 mL per bag, Rx only, Single Dose Injection, Refrigerated Injection, Central Admixture Pharmacy Services, Inc., … Incorrect Product Formulation |
| Aug 12 | Class I | Sunny Pharmtech Inc. Drug |
Cyclophosphamide for Injection, USP, 2 gram/vial, One Single-Dose Vial, Rx only, Manufactured for: Long Grove Pharmaceuticals, LLC, Rosemont, IL 60018. Manufactured in Taiwan, NDC 81298-8114-1 Presence of Particulate Matter; identified as stainless steel |
| Aug 12 | Class II | Aurobindo Pharma USA Inc Drug |
Dicyclomine Hydrochloride Capsules, USP, 10mg, 1,000 bottles, Rx only, Distributed by: Aurobindo Pharma USA, Inc., 279 Princeton-Hightstown Road, East Windsor, NJ 08520, Made in India, NDC 59651-719-99 shortfill; reports of empty capsules. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: Quantum 4Front 2HD Model/Catalog Number: Quantum 4Front 2HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: C… If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: Q6 Edge 2.0 (ilevel) Model/Catalog Number: Q6 Edge 2.0 (ilevel) Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Componen… If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | Stryker Corporation Device |
STRYKER PKG., STRYKEFLOW SUCTION/IRRIGATOR PS, Catalog Number 0250070540 There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: Q6 Edge 4 Model/Catalog Number: Q6 Edge 4 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | TISSIUM SA Device |
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. |
| Aug 12 | Class II | Stryker Corporation Device |
STRYKER AHTO TUBE SET WITH TIP PACKAGING, Catalog Number 0250070620 There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use. |
| Aug 12 | Class II | Maquet Critical Care AB Device |
Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30 6888530 Flow-i C40 6888540 Flow-c 6887700 Flow-e 6887900 Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness. |
| Aug 12 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; Model Numbers: 722003, 722010, 722026. 2. Allura Xper FD10 OR Table; Model Number: 722033. 3. Allura Xper… Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. |
| Aug 12 | Class II | United Orthopedic Corp. Device |
Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm Model/Catalog Number: 9323-5312 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A Due to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant. |
| Aug 12 | Class II | Boston Scientific Corporation Device |
Radial Jaw 4 Standard Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 SC 240cm Box 20; Outer Box UPN: M00513401; Inner Box UPN: M00513400. 2. Radial Jaw 4 SC 240cm Box 40; Oute… Potential for breach in sterile barrier of device packaging. |
| Aug 12 | Class II | TISSIUM SA Device |
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO6010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. |
| Aug 12 | Class II | Merge Healthcare, Inc. Device |
Merge OrthoPACS, medical imaging diagnostic workstation Thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open. |
| Aug 12 | Class II | INTEGRATED MEDCRAFT LLC Device |
Kingfisher Compact Individual First Aid Kit. REF Number: 76656. Containing the affected One-piece Guedel airway components: 1. Size 00, ISO 5.0, blue. Product Code: 1100050. 2. Size 000, ISO 3.5, light green. … Devices could potentially contain an occluded or partially occluded Guedel Airway tip. |
| Aug 12 | Class II | Aniara Diagnostica LLC Device |
5-PLASMA Bivalirudin Control, Model Number 5D-44905 Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values. |
| Aug 12 | Class II | TISSIUM SA Device |
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. |
| Aug 12 | Class II | Aniara Diagnostica LLC Device |
5-PLASMA Bivalirudin Calibrator, Model Number 5D-46905 Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values. |
| Aug 12 | Class II | Biomet, Inc. Device |
Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary Bone Saw Blade Assembly 14mm Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum rehab Product Name: Q6 Edge 2.0 Model/Catalog Number: Q6 Edge 2.0 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | Stryker Corporation Device |
STRYKER PKG., SINGLE SPIKE TUBE SET, Catalog Number 0250070530 There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use. |
| Aug 12 | Class III | Siemens Healthcare Diagnostics, Inc. Device |
Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated. |
| Aug 12 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; M… Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. |
| Aug 12 | Class II | Contego Medical, Inc Device |
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG… Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection. |
| Aug 12 | Class II | Instrumentation Laboratory Device |
GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for u… The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae. |
| Aug 12 | Class I | AVID Medical, Inc. Device |
Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. HEART CATH, SELF REGIONAL SELF131-05 Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels. |
| Aug 12 | Class II | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Device |
Philips Allura, Cardio Vascular-Allura Centron. Model Number: 722400. Interventional Fluoroscopic X-Ray System. Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives. |
| Aug 12 | Class II | Beckman Coulter Inc. Device |
REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284… Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications." |
| Aug 12 | Class II | Nuclein LLC Device |
Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH¿ Rapid PCR Instrument (DASH Instrument) is an automate… Due to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present). |
| Aug 12 | Class II | Therabody, Inc Device |
Charger for the Therabody Theragun, G4 PRO massage device, and products which include the 4th Generation Theragun PRO charger. G4 PRO Black US - Refurb., SKU TG02428-01 G4 PRO NA, SKU G4-PRO-PKG-NA G4-PRO-PKG-US… Charging cradle for the device may overheat while charging the device's removable battery pack, presenting risk of thermal event (fire). Firm has received reports alleging the product caught fire and other reports regarding visible smoke, melting, charger burn, burning odor, and overheating. |
| Aug 12 | Class II | Beckman Coulter Biomedical GmbH Device |
REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Model/Catalog Number: Q6 Edge 3 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | Medical Action Industries, Inc. 306 Device |
Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number: 78791 Convenience kits contain needle free access device that has an improper seal which may result in doming or tearing of the silicone seal. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: R-TRAK Model/Catalog Number: R-TRAK Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | Biomet, Inc. Device |
Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary Bone Saw Blade Assembly 16mm Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade. |
| Aug 12 | Class II | TISSIUM SA Device |
Brand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: N/A Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. |
| Aug 12 | Class II | Hardy Diagnostics Device |
HardyDisk AST Cefiderocol 30¿g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30¿g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enter… Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: Quantum Rival Model/Catalog Number: Quantum Rival Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | TISSIUM SA Device |
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. |
| Aug 12 | Class II | Biofire Defense Device |
FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the Fil… Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test. |
| Aug 12 | Class I | Boston Scientific Neuromodulation Corporation Device |
INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems. Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | Pride Mobility Products Corp. Device |
Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode. |
| Aug 12 | Class II | Biomet, Inc. Device |
Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475635 Product Description: Intramedullary Bone Saw Blade Assembly 15mm Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade. |
| Aug 12 | Class II | Merge Healthcare, Inc. Device |
Merge PACS, medical imaging diagnostic workstation Thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open. |
| Aug 12 | Class II | Biomet, Inc. Device |
Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone Saw Blade Assembly 12mm Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade. |
| Aug 12 | Class II | United Orthopedic Corp. Device |
Brand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptor, 2mm, 4mm, 6mm, left and Femoral valgus offset adaptor, 2mm, 4mm, 6mm, right Model/Catalog Number: 9323-5335, 9323-5337, 9323-5339 and 9323-5… Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail. |
| Aug 12 | Class II | MUNEVO GMBH Device |
munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders. |