5,412 recalls in the last 12 months. Newest first.

ReportedClassFirmProduct and reason
Jul 22 Class II GE Medical Systems, LLC
Device
Optima XR220amx, Mobile Digital-Ready Radiographic System
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Jul 22 Class II Daavlin Distributing Company
Device
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Jul 22 Class I Oscor Inc.
Device
Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
Jul 22 Class II Olympus Corporation of the Americas
Device
Olympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb. Model Number: 194.
Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
Jul 22 Class I Abiomed, Inc.
Device
ABIOMED Impella CP Introducer Kit 14 Fr, 13 and 25 cm, Pump Set Code 1000550, containing Kit, 14Fr Introducer, 13cm&25cm, Sterile, Introducer Product Code 0052-3046. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Jul 22 Class II GE Medical Systems, LLC
Device
Optima XR200amx, Mobile Digital-Ready Radiographic System
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Jul 22 Class II Daavlin Distributing Company
Device
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Jul 22 Class II ICU Medical, Inc.
Device
ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Jul 22 Class I Abiomed, Inc.
Device
ABIOMED Impella CP Introducer Kit 14 Fr, 13 cm, Pump Set Code 0052-0038, containing Kit, 14Fr Introducer, 13cm, Long Taper, Introducer Product Code 0052-3052. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Jul 22 Class II Papablic Inc
Device
Papablic Replacement Brush Heads. Model Number: Papablic18B
Potential for brush heads to break.
Jul 22 Class II Papablic Inc
Device
Papablic Replacement Brush Heads. Model Number: PAPA317AA.
Potential for brush heads to break.
Jul 22 Class I Abiomed, Inc.
Device
ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Jul 22 Class II Boston Scientific Corporation
Device
Rapid Refill Continuous Injection System, UPN M00566001
Devices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.
Jul 22 Class II Orthofix U.S. LLC
Device
Orthofix MINIRAIL, STANDARD LENGTHENER Model/Catalog Number: M101, Unilateral orthopedic/trauma external fixation device
Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
Jul 15 Class II Adams Extract & Spice LLC
Food
Johnny's Salad and Pasta Elegance 5.5 oz (UPC: 0-70381-10014-9) " Johnny's Salad and Pasta Elegance 2.75 oz (UPC: 0-70381-10013-2)
Undeclared Sesame allergen
Jul 15 Class I Fresh & Ready Foods LLC
Food
Fresh to You Burger Chicken Fillet Deluxe with Bacon, SKU 10303, Net wt. 5.9oz. Product is refrigerated. UPC 1 00000 01030 3. Manufactured by LPK1 Renton, WA. Product was not sold at retail stores and sold only to empl…
Raw breaded chicken patties were inadvertently used in the ready to eat Burger Chicken Deluxe With Bacon sandwich. Raw breaded chicken patties pose a risk of foodborne illness due to the potential presence of Salmonella.
Jul 15 Class I Lehi Valley Trading Company, Inc
Food
High Valley Orchard Chocolate Covered Raisins, net wt. 15oz. per package, each case contains 12/15oz. bags. Product is packaged in clear plastic container. UPC on retail container: 0 952348 80170 0. Lehi Valley Tradin…
Undeclared Peanut. High Valley Orchard Chocolate Covered Raisins contains unintended undeclared peanut due to mis-packaging.
Jul 15 Class I Adams Extract & Spice LLC
Food
JB Chicken Sknls 35.73#
Undeclared Sesame allergen
Jul 15 Class II Wondermade LLC
Food
Alabama Banana Pudding Marshmallow, 4 ounce 16 pieces of individual marshmallows packaged in a sealed plastic bag inside a printed cardboard box
Foreign material: small metal pieces
Jul 15 Class I Fresh & Ready Foods LLC
Food
Fresh to You Burger Chicken Fillet, SKU 10150, Net wt. 5oz. Product is refrigerated. UPC 1 00000 01015 0. Manufactured By LPK1 Renton, WA. Product was not sold at retail stores and sold only to employees at certain busi…
Raw breaded chicken patties were inadvertently used in the ready to eat Burger Chicken Fillet sandwich. Raw breaded chicken patties pose a risk of foodborne illness due to the potential presence of Salmonella.
Jul 15 Class I Adams Extract & Spice LLC
Food
Coopers Bbq 60 Gal Bean Mix box
Undeclared Sesame allergen
Jul 15 Class II Specialty Bakers, LLC
Food
Vanilla Dessert Shells (Giant), 6 dessert shells/1 package, 1 dessert shell is approximately 23.67 grams. 1package is 142 grams. 1 package has 1 plastic tray which holds 6 desserts shells. Plastic is wrapped around 1 tra…
Undeclared Allergen (Soy)
Jul 15 Class I Trafa Pharmaceutical Inc.
Food
Trafa Pharmaceuticals Inc. brand ORGANIC MORINGA LEAF POWDER; NET WEIGHT: 15 KGS; packaged in brown paper bags with a white label
Potential contamination with Salmonella
Jul 15 Class II Wise Foods Inc
Food
Wise Onion Rings, 3.75 oz, metalized film in cardboard cartons, 14 bags per case,
The product contains undeclared Yellow 5 and 6, Blue 1, and unapproved Red 3.
Jul 15 Class I Adams Extract & Spice LLC
Food
EC Hot Links Seasoning 45.05#
Undeclared Sesame allergen
Jul 15 Class II Bear Stewart Co
Food
Bakr Brown Butter Chocolate Chunk Ready to Bake Cookie Dough, 8-oz pouch 12 cookies per pouch, UPC 8 50073 08700 8, 8 pouches per case
undeclared soy
Jul 15 Class I Adams Extract & Spice LLC
Food
Smoked Sausage Spice 32.86#
Undeclared Sesame allergen
Jul 15 Class II QuVa Pharma, Inc.
Drug
Methohexital Sodium, 100 mg/10 mL (10 mg/mL), Total Volume: 10 mL, Rx only, Compounded drug, QuVa Pharma, 1075 W Park One Dr, Suite 100, Sugar Land, TX 77478, Product code: 70092-1310-46.
Failed Impurities/Degradation Specifications
Jul 15 Class II Gasco Industrial Corp.
Drug
Drogueria San Juan Puerto Rico Isopropyl Alcohol 70%, 16 oz, 474 mL, Distributed by: Drogueria San Juan, Calle De Diego 590, Sabana LLana Rio Piedras, PR 00924, UPC 69685696529
Subpotent drug
Jul 15 Class II BioMed Laboratories, LLC.
Drug
Medline Remedy Specialized Silicone Cream (Hydraguard-D ),Manufactured for Medline Industries, LP, Three Lakes Drive, Northfield, IL 80090 2 FL OZ (59 mL) NDC 53329-159-13 4 FL OZ (118 mL) NDC 53329-159-04
Failed Impurities/Degradation Specifications; Formaldehyde levels exceeded specification.
Jul 15 Class II Reliance Life Sciences Private Limited
Drug
Azacitidine for Injection 100mg/Vial, 1 10 mL Single-Dose Vial per carton, For Subcutaneous and Intravenous Use, Rx only, Manufactured for: Somerset Therapeutics, LLC., Somerset, NJ 08873 NDC: 70069-0857-01
Lack of Sterility Assurance
Jul 15 Class II Reliance Life Sciences Private Limited
Drug
Pemetrexed for Injection 100mg/Vial, 1 10 mL Single-Dose Vial per carton, For intravenous use only, Rx only, Manufactured for: Somerset Therapeutics, LLC., Somerset, NJ 08873 NDC: 70069-0834-01
Lack of Sterility Assurance
Jul 15 Class II Reliance Life Sciences Private Limited
Drug
Bortezomib for Injection 3.5mg/Vial, 10 mL per Single-Dose Vial, For Intravenous or Subcutaneous Use, Manufactured for: Somerset Therapeutics, LLC., Somerset, NJ 08873. NDC: 70069-0836-01
Lack of Sterility Assurance
Jul 15 Class II Gasco Industrial Corp.
Drug
Drogueria San Juan Puerto Rico Isopropyl Alcohol 70%, 1 gallon (3.78 liter), Distributed by: Drogueria San Juan, Calle De Diego 590, Sabana LLana Rio Piedras, PR 00924
Subpotent drug
Jul 15 Class II Gasco Industrial Corp.
Drug
Gasco Isopropyl Alcohol 70%, Net Contents: 16 Oz. (474 mL). Manufactured by: Gasco Industrial Corp., PO Box 1360 Gurabo PR 00778 Made in USA UPC 7 87302 13044 3
Subpotent drug
Jul 15 Class II Reliance Life Sciences Private Limited
Drug
Pemetrexed for Injection 500mg/Vial, 1 50 mL Single-Dose Vial per carton, For intravenous use only, Rx only, Manufactured for: Somerset Therapeutics, LLC., Somerset, NJ 08873 NDC: 70069-0835-01
Lack of Sterility Assurance
Jul 15 Class II Shimoga Chemicals
Drug
Clomiphene Citrate USP Active Pharmaceutical Ingredient (CAS : 50-41-9), Net Weight a)1 KG (NDC 84849-000-01), b) 500 g (NDC 84849-000-02), (c) 100 g (NDC 84849-000-03), Rx Only, Shimoga Chemicals Address -W-57A, MIDC Ku…
cGMP deviations
Jul 15 Class II Gasco Industrial Corp.
Drug
Drogueria San Juan Puerto Rico Isopropyl Alcohol 70%, 32 oz., 946 mL, Distributed by: Drogueria San Juan, Calle De Diego 590, Sabana Llana Rio Piedras, PR 00924, UPC 659685696567.
Subpotent drug
Jul 15 Class II Gasco Industrial Corp.
Drug
Gasco Isopropyl Alcohol 70%, 32 oz (946 mL) Gasco Industrial Corp., PO Box 1360, Gurabo PR 00778, UPC 7 87302 12363 6.
Subpotent drug
Jul 15 Class II Gasco Industrial Corp.
Drug
Gasco Isopropyl Alcohol 70%, Net Contents: 1 US Gallon (3.78 L), Manufactured by: Gasco Industrial Corp., PO Box 1360, Gurabo, PR 00778, Made in USA
Subpotent drug
Jul 15 Class II ONESOURCE SPECIALTY PHARMA LIMITED
Drug
Methohexital Sodium for Injection, USP 500 mg Multiple Dose Vial, Rx Only, Manufactured by: OneSource Specialty, Pharma Limited, Bengaluru - 561 203, India, Manufactured for: Avet Pharmaceuticals Inc., East Brunswick, NJ…
Failed Impurities/Degradation Specifications
Jul 15 Class II Integra LifeSciences Corp. (NeuroSciences)
Device
OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Jul 15 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) OPEN HEART PEDS PACK, Model Number: DYNJ88569D; 2) ACCESSORY KIT, Model Number: DYNJ901348F
Convenience Kits contain Covidien Mon-a-Therm General Purpose Temperature Probes Size 9 Fr. Medtronic has issued a recall of specific production lots of the Covidien Mon-a-Therm general purpose temperature probe-size 9 due to an error in its distribution control process.
Jul 15 Class II Encore Medical, LP
Device
Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 341-14-711 Product Description: Enovis knee devices are for knee arthroplasty.
Due to packaging discrepancy with incorrect product size mixed up.
Jul 15 Class II Encore Medical, LP
Device
Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 342-19-703 Product Description: Enovis knee devices are intended for knee arthroplasty.
Due to packaging discrepancy with incorrect product size mixed up.
Jul 15 Class II Medline Industries, LP
Device
Medline Convenience Kits: 1) OPEN HEART PEDS PACK, Model Number: DYNJ88569; 2) OPEN HEART PEDS PACK, Model Number: DYNJ88569A; 3) OPEN HEART PEDS PACK, Model Number: DYNJ88569B; 4) OPEN HEART PEDS PACK, Model Numb…
Convenience Kits contain Covidien Mon-a-Therm General Purpose Temperature Probes Size 9 Fr. Medtronic has issued a recall of specific production lots of the Covidien Mon-a-Therm general purpose temperature probe-size 9 due to an error in its distribution control process.
Jul 15 Class II Boston Scientific Neuromodulation Corporation
Device
Intracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DRILL/RLV0035, INTRACEPT ACCESS INSTRUMENTS- DRILL EU/RLV0035-EU, INTRACEPT ACCESS INSTRUMENTS- 2VB/RLV0045, INTRACEPT ACCESS INSTRUMENTS- …
Intraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility breach, and if breached could lead to severe infection.
Jul 15 Class II Biomerieux Inc
Device
VITEK 2 AST-N447 TEST KIT Product reference 424620
Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.
Jul 15 Class II Think Surgical, Inc.
Device
TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system fo…
Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
Jul 15 Class II Elekta, Inc.
Device
Brand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended for intracavitary brachytherapy for cancer treatment of the cervix and the endometrium. Optional needles can be plac…
Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.