FDA Brief

Recalls

When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.

5,412 recalls in the last 12 months. Newest first.

ReportedClassFirmProduct and reason
Dec 31, 2025 Class II Cook Incorporated
Device
Ring Transjugular Intrahepatic Access Sets, Reference Part Numbers: 1) RTPS-100; 2) RTPS-100-10.0; The Ring Transjugular Intrahepatic Access Set is intended for transjugular liver access in diagnostic and interven…
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
FDA notice Z-0994-2026
Dec 31, 2025 Class II Beckman Coulter Ireland, Inc.
Device
BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the quantitative determination of Bicarbonate in human serum and plasma on Beckman Coulter AU analyzers.
Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.
FDA notice Z-0997-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609360 Software Version: N/A Product Description: Affixus 9 mm, Length 360 mm, Left, Trochanteri…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0977-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509320 Software Version: N/A Product Description: Affixus 9 mm, Length 320 mm, Right, Trochanter…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0967-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609380 Software Version: N/A Product Description: Affixus 9 mm, Length 380 mm, Left, Trochanteri…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0978-2026
Dec 31, 2025 Class II Diasol, Inc
Device
Brand Name: Diasol Product Name: 100325-10-DEX100, Liquid Concentrate for Bicarbonate Dialysis Non Pyrogenic, 2K, 3.0CA, 1.0MG, 208.2 liters
Safety and efficacy of dialysis acid concentrate cannot be assured
FDA notice Z-0991-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509300 Software Version: N/A Product Description: Affixus 9 mm, Length 300 mm, Right, Trochanter…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0983-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609340 Software Version: N/A Product Description: Affixus 9 mm, Length 340 mm, Left, Trochanteri…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0976-2026
Dec 31, 2025 Class II Cook Incorporated
Device
R¿sch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The R¿sch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
FDA notice Z-0995-2026
Dec 31, 2025 Class II Microbiologics Inc
Device
LYFO DISK, Campylobacter jejuni subsp. jejuni derived from ATCC 29428, Catalog Number 0325L
Affected units may fail to recover the target microorganism Campylobacter jejuni subsp. jejuni derived from ATCC 29428.
FDA notice Z-0986-2026
Dec 31, 2025 Class II Microbiologics Inc
Device
KWIK-STIK 6-Pack, Campylobacter jejuni subsp. jejuni derived from ATCC 29428, Catalog Number 0325K
Affected units may fail to recover the target microorganism Campylobacter jejuni subsp. jejuni derived from ATCC 29428.
FDA notice Z-0985-2026
Dec 31, 2025 Class II AGFA Healthcare Corp.
Device
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
FDA notice Z-0923-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609420 Software Version: N/A Product Description: Affixus 9 mm, Length 420 mm, Left, Trochanteri…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0980-2026
Dec 31, 2025 Class II ICU Medical, Inc.
Device
LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
FDA notice Z-0966-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509360 Software Version: N/A Product Description: Affixus 9 mm, Length 360 mm, Right, Trochanter…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0969-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509400 Software Version: N/A Product Description: Affixus 9 mm, Length 400 mm, Right, Trochanter…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0971-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609300 Software Version: N/A Product Description: Affixus 9 mm, Length 300 mm, Left, Trochanteri…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0974-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815809280 Software Version: N/A Product Description: Affixus 9 mm, Length 280 mm, Left, Priformis N…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0982-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509340 Software Version: N/A Product Description: Affixus 9 mm, Length 340 mm, Right, Trochanter…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0968-2026
Dec 31, 2025 Class II Intuitive Surgical, Inc.
Device
da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
FDA notice Z-0964-2026
Dec 31, 2025 Class II Diasol, Inc
Device
Brand Name: Diasol Product Name: 100220-10-DEX100, Liquid Concentrate for Bicarbonate Dialysis Non Pyrogenic, 2K, 3.0CA, 1.0MG, 208.2 liters
Safety and efficacy of dialysis acid concentrate cannot be assured
FDA notice Z-0989-2026
Dec 31, 2025 Class II Diasol, Inc
Device
Brand Name: Diasol Product Name: 100225-10-DEX100, Liquid Concentrate for Bicarbonate Dialysis Non Pyrogenic, 2K, 3.0CA, 1.0MG, 208.2 liters
Safety and efficacy of dialysis acid concentrate cannot be assured
FDA notice Z-0987-2026
Dec 31, 2025 Class II Diasol, Inc
Device
Brand Name: Diasol Product Name: 100425-10-DEX100, Liquid Concentrate for Bicarbonate Dialysis Non Pyrogenic, 2K, 3.0CA, 1.0MG, 208.2 liters
Safety and efficacy of dialysis acid concentrate cannot be assured
FDA notice Z-0992-2026
Dec 31, 2025 Class II Microbiologics Inc
Device
KWIK-STIK 2-Pack, Campylobacter jejuni subsp. jejuni derived from ATCC 29428, Catalog Number 0325P
Affected units may fail to recover the target microorganism Campylobacter jejuni subsp. jejuni derived from ATCC 29428.
FDA notice Z-0984-2026
Dec 31, 2025 Class II Cook Incorporated
Device
Flexor Check-Flo Introducers and Sets, Reference Part Numbers: 1) KCFW-9.0-18/38-45-RB-ANL0-HC; 2) KCFW-9.0-18/38-45-RB-ANL1-HC; 3) KCFW-10.0-35-45-RB-HFANL0-HC; 4) KCFW-10.0-35-45-RB-HFANL1-HC; 5) KCFW-12.0-3…
Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.
FDA notice Z-0993-2026
Dec 31, 2025 Class I Draeger, Inc.
Device
Draeger Vapor 2000. Unheated, calibrated anesthetic vaporizer.
A certain component of affected devices was not delivered within specification and contained impurities.
FDA notice Z-0933-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609320 Software Version: N/A Product Description: Affixus 9 mm, Length 320 mm, Left, Trochanteri…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0975-2026
Dec 31, 2025 Class II Howmedica Osteonics Corp.
Device
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which inc…
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
FDA notice Z-0963-2026
Dec 31, 2025 Class II ICU Medical, Inc.
Device
LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02
For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.
FDA notice Z-0965-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509380 Software Version: N/A Product Description: Affixus 9 mm, Length 380 mm, Right, Trochanter…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0970-2026
Dec 31, 2025 Class II Zimmer, Inc.
Device
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509420 Software Version: N/A Product Description: Affixus 9 mm, Length 420 mm, Right, Trochanter…
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
FDA notice Z-0972-2026
Dec 31, 2025 Class II Medtronic Neuromodulation
Device
Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TAB…
Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.
FDA notice Z-0996-2026
Dec 31, 2025 Class II CareFusion 303, Inc.
Device
BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with…
Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.
FDA notice Z-0962-2026
Dec 24, 2025 Class II Hollandia Dairy, Inc.
Food
Hollandia Dairy, Inc. 100% Appleberry Juice; Half Pint (236mL). UPC: 0 27495 03679 6
100% Appleberry Juice contaminated with cleaning solution, Peracetic Acid.
FDA notice H-0307-2026
Dec 24, 2025 Class I Lexunder Inc
Food
Food to Live brand ORGANIC SUPERGREENS POWDER MIX, packaged heat-sealed, resealable stand-up pouches in 8 oz, 1 lb., 1.5 lb., 3 lb., 6 lb., 12 lb.; Ingredients: ORGANIC KALE; ORGANIC MORINGA LEAF; ORGANIC WHEAT GRASS; OR…
Contaminated with Salmonella
FDA notice H-0310-2026
Dec 24, 2025 Class II Shata Traders Inc.
Food
Chef brand Milk Pan 24cm
Products found to contain leachable lead levels (3.00 mg/L)
FDA notice H-0306-2026
Dec 24, 2025 Class II Good Way Trading Group Inc
Food
BaiLiFeng Dried Cinnamon Powder; Net Wt 4 oz/114G; Good Way Trading Group Inc 58-51 Maspeth Avenue Maspeth NY 11378; Product of China; Ingredients: Dried Cinnamon; UPC code: 4 897055 79940 1
contains high levels of lead (10.7 mg/kg)
FDA notice H-0313-2026
Dec 24, 2025 Class I MAMTAKIM, Inc.
Food
Belevini KORUSHKA, Dried Salted Whole Smelts, Net. Weight, 5.29 oz (150 g) VACUUM PACK, 20 PACKAGES PER CASE, FROZEN, REFRIGERATE AFTER OPENING
Clostridium botulinum (uneviscerated fish)
FDA notice H-0311-2026
Dec 24, 2025 Class I Lexunder Inc
Food
Food to Live brand ORGANIC MORINGA LEAF POWDER, packaged heat-sealed, resealable stand-up pouches in 8 oz, 1 lb., 2 lb., 4 lb., 8 lb., 16 lb., and 44 lb. bags; Ingredients: ORGANIC MORINGA LEAF POWDER; UPC: 8 oz - 840…
Contaminated with Salmonella
FDA notice H-0309-2026
Dec 24, 2025 Class II Dough-To-Go Inc
Food
Linzer Hearts Baked Cut Out Cookies UPC 0 44964 22072 2 72 - 4.0 oz Cut Cookies CBB-220-LZ Net Weight: 18 lbs Ingredients: Enriched flour (bleached WHEAT flowr, malted barley flour, niacin, iron, thaimine mononitrate, ri…
Undeclared Milk and FD&C Blue #1.
FDA notice H-0364-2026
Dec 24, 2025 Class I MAMTAKIM, Inc.
Food
Belevini BICHKI, Dried Salted Whole Smelts, Net. Weight, 5.29 oz (150 g) VACUUM PACK,20 PACKAGES PER CASE, FROZEN, REFRIGERATE AFTER OPENING
Clostridium botulinum (uneviscerated fish)
FDA notice H-0312-2026
Dec 24, 2025 Class II The New York Milk Bank, Inc.
Food
Pasteurized Donor Human Milk in 4 ounce/120 ml clear plastic bottles. The bottles have a foil seal and are capped with a green lid. Bottle size: 120ml
Foreign object
FDA notice H-0308-2026
Dec 24, 2025 Class I The Whole Herb Company, Inc.
Food
SENNA LEAVES F/C OG Item No: SEN15001 Net Wt: 44.75 Lbs Distributed By Whole Herb Company 19800 8th Street East, Sonoma, CA 95476 UPC 10031 44312
Potential contamination with Salmonella.
FDA notice H-0325-2026
Dec 24, 2025 Class II CELEBRATION HERBALS US
Food
Celebration Herbals Senna Leaf Herbal Tea 24 Teabags, net wt. 1.48oz. UPC 628240201829
possible salmonella
FDA notice H-0328-2026
Dec 24, 2025 Class II Novocol Pharmaceutical of Canada, Inc.
Drug
Lidocaine HCI 2% and Epinephrine 1:100,000 (Lidocaine HCI and Epinephrine Injection USP), packaged in a carton containing 50 single-dose Cartridges, 1.7 ml each, Rx only, Manufactured for Patterson Dental Supply, Inc., 1…
Defective container: cracked/broken cartridges
FDA notice D-0229-2026
Dec 24, 2025 Class II Novocol Pharmaceutical of Canada, Inc.
Drug
Cook-Walte, Carbocaine 3% (30 mg/mL), (mepivacaine hydrochloride Injection, USP), packaged in a carton containing 50 Single-Dose Cartridges, 1.7 mL, Rx only, Manufactured by Novocol Pharmaceutical of Canada, Inc., NDC 0…
Defective container: cracked/broken cartridges
FDA notice D-0233-2026
Dec 24, 2025 Class III SOMERSET THERAPEUTICS LLC
Drug
Cisatracurium Besylate Injection USP, 10 mg/5 mL* (2mg/mL), 10 x 5 mL Single-dose Vials per Carton, Rx Only, Manufactured for: Somerset Therapeutics, LLC, Hollywood, FL 33024, Made in India, Single-dose 5 mL vial NDC 700…
Subpotent product:out of specification assay results observed during long term stability testing.
FDA notice D-0236-2026
Dec 24, 2025 Class II Novocol Pharmaceutical of Canada, Inc.
Drug
OraVerse, (Phentolamine Mesylate) Injection, 0.4 mg/1.7mL, packaged in a carton containing 10 cartridges, 0.4 mg/1.7mL, Rx only, Made in Canada by Novocol Pharmaceutical of Canada, Inc., Distributed by Septodont, Inc., L…
Defective container: cracked/broken cartridges
FDA notice D-0234-2026
Dec 24, 2025 Class II Novocol Pharmaceutical of Canada, Inc.
Drug
2% Xylocaine DENTAL with Epinephrine 1:100,000 (Lidocaine HCI and Epinephrine Injection USP), packaged in a carton containing 50 single-dose cartridges, 1.7 mL each, Rx only, Manufactured for Dentsply Pharmaceutical, Yor…
Defective container: cracked/broken cartridges
FDA notice D-0228-2026
Dec 24, 2025 Class III SOMERSET THERAPEUTICS LLC
Drug
Cisatracurium Besylate Injection USP, 200mg/20mL*(10mg/mL), 10 x 20 mL Single-dose Vials per Carton, Rx Only, Manufactured for: Somerset Therapeutics, LLC, Hollywood, FL 33024, Made in India, Single dose 20mL vial NDC 70…
Subpotent product:out of specification assay results observed during long term stability testing.
FDA notice D-0237-2026