Recalls
When a product is unsafe or breaks the rules, the company pulls it back and FDA classes the recall. Class I could cause serious harm or death. Class II could cause temporary harm. Class III is unlikely to hurt anyone.
5,412 recalls in the last 12 months. Newest first.
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340. The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors. FDA notice Z-1612-2026 |
| Apr 1 | Class II | Tandem Diabetes Care, Inc. Device |
Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion … The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event. FDA notice Z-1636-2026 |
| Apr 1 | Class I | Merit Medical Systems, Inc. Device |
16F Dual Valved Splittable Sheath Introducer (bulk, non-sterile), REF: FCL-174-00/B 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access. FDA notice Z-1577-2026 |
| Apr 1 | Class II | Waldemar Link GmbH & Co. KG (Mfg Site) Device |
Embrace Drill Tower, Wedged (25mm); Item Number: 645-081/62; The surgical technique for the Embrace Shoulder Instruments - Drill Tower has mixed-up article numbers in the overview of the instruments. However, the individual surgical steps described in the surgical technique are correct. FDA notice Z-1651-2026 |
| Apr 1 | Class II | Medtronic MiniMed, Inc. Device |
MiniMed" 780G - MMT-1884 MiniMed" 780G - MMT-1886 With Software 6.60 and 6.61 Three software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due to inaccuracies in insulin delivery (either under-delivery, over-delivery, or cessation of delivery). FDA notice Z-1635-2026 |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alterna… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. FDA notice Z-1640-2026 |
| Apr 1 | Class III | Philips Ultrasound, Inc Device |
Philips OMNI II TEE Ultrasound Transducer To provide clarification and labelling to define the useful life of ultrasound transducers in the field. FDA notice Z-1623-2026 |
| Apr 1 | Class III | Philips Ultrasound, Inc Device |
Philips X3-1 Ultrasound Transducer To provide clarification and labelling to define the useful life of ultrasound transducers in the field. FDA notice Z-1627-2026 |
| Apr 1 | Class II | CareFusion 303, Inc. Device |
BD Alaris" System with Guardrails" Suite MX with Point of Care Unit Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units FDA notice Z-1645-2026 |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140. The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors. FDA notice Z-1613-2026 |
| Apr 1 | Class I | Philips Respironics, Inc. Device |
Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device. Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient. FDA notice Z-1615-2026 |
| Apr 1 | Class II | GE Medical Systems, LLC Device |
Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position. FDA notice Z-1590-2026 |
| Apr 1 | Class III | Philips Ultrasound, Inc Device |
Philips S4-1 Ultrasound Transducer To provide clarification and labelling to define the useful life of ultrasound transducers in the field. FDA notice Z-1625-2026 |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus PKS Cutting Forceps w/Cord, 5mm 24cm Product Name: PKS Cutting Forceps Model/Catalog Number: 920000PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. C… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. FDA notice Z-1643-2026 |
| Apr 1 | Class I | Merit Medical Systems, Inc. Device |
ProGuide Chronic Dialysis Catheters, REF: DC01455550/C, DC21452419/C, DC21452419-NE5/C, DC21452823-NE5/C, DC21454035/C, DC21455550/C 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access. FDA notice Z-1576-2026 |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443. The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors. FDA notice Z-1610-2026 |
| Apr 1 | Class II | I.T.S. GmbH Device |
Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column P… Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). FDA notice Z-1608-2026 |
| Apr 1 | Class II | I.T.S. GmbH Device |
I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers… Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). FDA notice Z-1598-2026 |
| Apr 1 | Class I | Erbe USA Inc Device |
Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use Probes may rupture/burst during activation FDA notice Z-1568-2026 |
| Apr 1 | Class I | Merit Medical Systems, Inc. Device |
CentrosFLO Hemodialysis Catheters, REF: CENFP15K/A, CENFP15K/B, CENFP17K/A, CENFP19K/A, CENFP23K/A, CENFP27K/A, CENFP31K/A, CENFT15K, CENFT15K/D, CENFT17K, CENFT17K/A, CENFT17K/D, CENFT19K, CENFT19K/A, CENFT19K/B, CENFT1… 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access. FDA notice Z-1575-2026 |
| Apr 1 | Class II | I.T.S. GmbH Device |
Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 373… Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). FDA notice Z-1602-2026 |
| Apr 1 | Class II | I.T.S. GmbH Device |
HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-1… Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). FDA notice Z-1607-2026 |
| Apr 1 | Class II | Olympus Corporation of the Americas Device |
Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulatio… Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. FDA notice Z-1642-2026 |
| Apr 1 | Class II | I.T.S. GmbH Device |
I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 314… Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). FDA notice Z-1599-2026 |
| Apr 1 | Class III | Philips Ultrasound, Inc Device |
Philips S5-2 Ultrasound Transducer To provide clarification and labelling to define the useful life of ultrasound transducers in the field. FDA notice Z-1626-2026 |
| Apr 1 | Class II | I.T.S. GmbH Device |
Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6… Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). FDA notice Z-1597-2026 |
| Apr 1 | Class I | Erbe USA Inc Device |
Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use Probes may rupture/burst during activation FDA notice Z-1566-2026 |
| Apr 1 | Class II | I.T.S. GmbH Device |
FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Nu… Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). FDA notice Z-1605-2026 |
| Apr 1 | Class I | Philips Respironics, Inc. Device |
Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device. Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient. FDA notice Z-1616-2026 |
| Apr 1 | Class I | Philips Respironics, Inc. Device |
Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device. Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient. FDA notice Z-1617-2026 |
| Apr 1 | Class II | I.T.S. GmbH Device |
HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31… Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). FDA notice Z-1606-2026 |
| Apr 1 | Class II | Orthalign, Inc Device |
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm) FDA notice Z-1630-2026 |
| Apr 1 | Class II | Orthalign, Inc Device |
Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm) FDA notice Z-1629-2026 |
| Mar 25 | Class II | ASUKAR FOODS CORP Food |
TETAS Mireya, Cherry Milkshake (Colita con Leche), i unit Popsicle, Net wt 6OZ. 170GRS. KEEP FROZEN. UPC Code 83822654822 Undeclared allergen: Soy Lecithin FDA notice H-0584-2026 |
| Mar 25 | Class II | Wawa Beverage Company Food |
Wawa Cookies & Cream Milk 16 oz (Pint), Plastic bottle Store Refrigerated Foreign plastic material discovered on fill line. FDA notice H-0587-2026 |
| Mar 25 | Class I | Bluegrass Ingredients Inc. Food |
Butter Powder, Net Weight 50 LB., packaged in a multiply Kraft bag with poly liner Positive Salmonella contamination FDA notice H-0573-2026 |
| Mar 25 | Class II | SAPUTO CHEESE USA, INC Food |
CLOVER SONOMA SOUR CREAM NET WT 8OZ UPC 0 70852 61800 5 & 16OZ UPC 0 70852 61500 4 DISTRIBUTED BY: CLOVER-STORNETTA FARMS LLC PETALUMA, CA 94952 Potentially under pasteurized. The firm was notified by the California Department of Agriculture that due to the broken seal on the pasteurizer, the products may be under pasteurized, and they should recall the products. FDA notice H-0580-2026 |
| Mar 25 | Class II | BlueFrog Foods Food |
GREENWAY FARMS of Georgia GARLIC DILL PICKLES CHIPS Net Wt 16 fl oz (473 ml) UPC 8 57866 00501 0 Produced by: Greenway Farms of Georgia, LLC Roberta, GA Packed by Blue Frog Food, LLC for Green way Farms of Georgia, LLC 5… No process authority FDA notice H-0577-2026 |
| Mar 25 | Class II | Wawa Beverage Company Food |
Wawa Chocolate Low-fat Milk 16 oz , (Pint) Plastic bottle .Store Refrigerated Foreign plastic material discovered on fill line. FDA notice H-0588-2026 |
| Mar 25 | Class I | Shaman Botanicals, LLC Food |
Product label reads in part as "WHITE VEIN ADVANCED ALKALOIDS CHEWABLE TABLETS 15mg 7-Hydroxymitragynine per container (7.5mg 7-OH per serving) LEMON DIETARY SUPPLEMENT Net Wt. 0.02 oz 2 TABLETS Supplement Facts 7-hydrox… A sample of the product had the active ingredient in a quantity greater than the labeled specification. FDA notice H-1107-2026 |
| Mar 25 | Class I | Shaman Botanicals, LLC Food |
Product label reads in part as "WHITE VEIN ADVANCED ALKALOIDS CHEWABLE TABLETS 225mg 7-Hydroxymitragynine per container (7.5mg 7-OH per serving) LEMON DIETARY SUPPLEMENT Net Wt. 0.25 oz 30 TABLETS Supplement Facts 7-hydr… A sample of the product revealed the active ingredient is in a quantity greater than the labeled specification. FDA notice H-1108-2026 |
| Mar 25 | Class II | Dr. Praeger's Sensible Foods Inc. Food |
Simply Nature, Spinach Bites, 12 oz retail unit, Plastic wrapped bites inside of a retail box. Keep Frozen. This product has the possibility of contamination with rodent hair. FDA notice H-0570-2026 |
| Mar 25 | Class I | FRITO-LAY NORTH AMERICA, INC. Food |
Miss Vickies Spicy Dill Pickle; 8 oz bag Undeclared Allergen: Milk FDA notice H-0571-2026 |
| Mar 25 | Class II | SAPUTO CHEESE USA, INC Food |
CLOVER SONOMA COTTAGE CHEESE LOW FAT 2% MILKFAT NET WT 16OZ UPC 0 70852 54400 7 & 32OZ UPC 0 70852 54200 3 DISTRIBUTED BY: CLOVER-STORNETTA FARMS LLC PETALUMA, CA 94952 Potentially under pasteurized. The firm was notified by the California Department of Agriculture that due to the broken seal on the pasteurizer, the products may be under pasteurized, and they should recall the products. FDA notice H-0579-2026 |
| Mar 25 | Class I | Bluegrass Ingredients Inc. Food |
Butter-Rich Powder, Net Weight 50 LB., packaged in a multiply Kraft bag with poly liner Potential Salmonella contamination FDA notice H-0572-2026 |
| Mar 25 | Class I | Kradjian Importing Company, Inc. Food |
MID EAST Jordanian Zaatar spice mix; NET Wt. 5lb plastic bag; UPC: 094379607038 Packed by: Propack Inc. Sunvalley, CA 91354 Undeclared allergen ingredients (sesame and wheat) in Zaatar Spice mix. FDA notice H-0576-2026 |
| Mar 25 | Class I | Dreamland Inc Food |
Dreamland's Zaatar Chickpea Salad is packaged in poly bags with a generic label which declares "chickpeas, cumin, zataar, turmeric, dill, onion, garlic, parsley". Dreamland distributed bulk cases (2/5lbs. bags per case) … Undeclared Sesame, Soy, Wheat FDA notice H-0575-2026 |
| Mar 25 | Class II | Wawa Beverage Company Food |
Wawa Double Dutch Chocolate Milk ,16 oz (Pint) Plastic bottle Store Refrigerated Foreign plastic material discovered on fill line. FDA notice H-0586-2026 |
| Mar 25 | Class III | Winder Laboratories, LLC Food |
Multivitamin with Fluoride Chewable Tablets Grape Flavor Rx NDC 75826-169-10 0.25 mg 100 TABLETS UPC 3 15826 16910 2; NDC 75826-171-10 1mg 100 TABLETS UPC 3 15826 17110 5 Manufactured by: WINDER LABORATORIES LLC Winder, … Subpotent. During an FDA inspection it was discovered that the product was subpotent for Vitamin D. FDA notice H-0569-2026 |
| Mar 25 | Class II | SAPUTO CHEESE USA, INC Food |
Great Value Cottage Cheese Small Curd 4% Milkfat NET WT 16oZ UPC 0 78742 37235 8; 24oz UPC 0 78742 37236 5 & 3lb UPC 0 78742 14797 0 DISTRIBUTED BY: Walmart Inc., Bentonville, AR 72716 Potentially under pasteurized. The firm was notified by the California Department of Agriculture that due to the broken seal on the pasteurizer, the products may be under pasteurized, and they should recall the products. FDA notice H-0583-2026 |